Description
Bair Hugger 750 Patient Warming System – Refurbished. The Bair Hugger 750 is designed for accurate, efficient temperature management. Hose-end temperature sensing ensures precise air temperature delivery, while three integrated temperature sensors guarantee rapid response. The unit’s increased airflow further enhances quick and effective warming, making it ideal for diverse clinical needs.
This advanced system tracks key therapy metrics, including duration, over-temperature events, and calibration status. Additionally, the built-in hour meter records usage, and fault codes are easily accessible via the front panel. There’s no need to open the unit to access this essential information, streamlining operation and maintenance.
The Bair Hugger 750 directly integrates with the Bair Hugger 241® blood/fluid warming set, making it the only forced-air warming system to do so. Moreover, this unique compatibility offers a cost-effective method for fluid warming, making the Bair Hugger 750 a versatile and efficient solution for comprehensive patient care.
- See tabs for addition information on the Bair Hugger 750.
Dimensions:
- Size: 12.5” H x 13.5” D x 10” ,W 31.5 cm H x 34 D x 25 cm W
- Weight: 16.3 lb (7.4 kg)
Relative Noise Level:
Hose:
- Flexible, compatible with Bair Hugger brand 241 blood/fluid warming system.
Filtration System:
- High efficiency 0.2 μm Filter
Recommended Filter change:
- Every 12 months or 500 hours of use
Mounting:
- Can be clamped to an IV pole, placed on a hard surface, or mounted to the rolling stand accessory.
Recommended Operating Environment:
- Temperature: 15°C-25°C
- Humidity: Max relative humidity 80% (up to 31°C) decreasing linearly to 50% relative humidity at 40°C
- Altitude: Max 2,000m
Temperature Control:
- Electronically Controlled
Heat Generated:
- 1,644 BTU/hr (average), 482 W (average)
Operating Temperatures:
- Average Temperature at the end of the hose:
- High: 43°±1.5 °C 109.4°± 2.7°FMed: 38°±1.5 °C 100.4°± 2.7°F
Time to reach Operating Temperature:
- 2 – 5 minutes (dependent on blanket model)
- Time required for the contact surface temperature to heat up from 23 ± 2°C to 37°C (73 ± 2°F to 99°F)
Storage/Transport Temperature:
- -20°C to 45°C (-4°F to 113°F)
- Store all components in a cool, dry place when not in use.